Performance characteristics of high-frequency percussive ventilation under hyperbaric conditions
Introduction: Safe administration of critical care hyperbaric medicine requires specialized equipment and advanced training. Equipment must be tested in order to evaluate function in the hyperbaric environment. High-frequency percussive ventilation (HFPV) has been used in intensive care settings effectively, but it has never been tested in a hyperbaric chamber.
Methods: Following a modified U.S. Navy testing protocol used to evaluate hyperbaric ventilators, we evaluated an HFPV transport ventilator in a multiplace hyperbaric chamber at 1.0, 1.9, and 2.8 atmospheres absolute (ATA). We used a test lung with analytical software for data collection. The ventilator uses simultaneous cyclic pressure-controlled ventilation at a pulsatile flow rate (PFR)/oscillatory continuous positive airway pressure (oCPAP) ratio of 30/10 with a high-frequency oscillation percussive rate of 500 beats per minute. Inspiratory and expiratory times were maintained at two seconds throughout each breathing cycle.
Conclusion: The HFPV transport ventilator performed adequately under hyperbaric conditions and should be considered a viable option for hyperbaric critical care. This ventilator has atypical terminology and produces unique pulmonary physiology, thus requiring specialized training prior to use.